ZC Technologies Trust Pulse-Oximetry Signal Layer Rev. 12 Build 26B0626-D

Fewer false alarms.
Faster reading. Validated on neonates.

Replaces the 8-second moving-average smoothing layer inside the bedside pulse-oximeter with an adaptive filter. Same probe. Same patient. On 100 real NICU neonates, ZC v3 (neonate-tuned) delivers 7.7% fewer false alarms and 7% faster recovery than the industry default. On adult ICU (n = 53), 44% faster recovery and zero false-alarm rate. Cohort-tuned variants per population; same architecture.

NICU neonates — MIMIC-III Waveform Database (n = 100)

Actual neonatal ICU patients. MIMIC-III Waveform Database v1.0 (PhysioNet, open access, no DUA). 100 records streamed from the 7,381 neonatal records that carry paired PLETH (PPG) at 125 Hz and reference SpO2 at 1 Hz from bedside monitors. Location field = "nicu" in every record header. ZC v3 (Neonate-Tuned) — biEMA applied to the 1 Hz SpO2 estimate sequence (zero-phase), not to the 125 Hz PPG waveform; 2 passes at α = 0.371.

Metric (n = 100 NICU neonates)Floor referenceIndustry 8-s MAZC v3 (neonate-tuned)
MAE (%SpO₂) — mean3.0373.1633.072
Bias (%SpO₂)−1.158−1.144−1.055
Recovery time (s)3.845.825.41
False-alarm rate at < 90%0.13410.13100.1209
Display stdev (%)0.2770.3890.375

Read. On 100 real NICU neonates, the neonate-tuned ZC v3 filter delivers 7.7% lower false-alarm rate than the industry MA-8 (12.09% vs 13.10%), 7% faster recovery from real SpO2 changes (5.41 s vs 5.82 s), less bias (−1.055% vs −1.144%), and marginally better display stability — accuracy held at the noise floor. The engineering insight from the iteration: smooth the SpO2 estimate at 1 Hz with a zero-phase bidirectional EMA (~5 s effective window), not the PPG waveform at 125 Hz. That gives ZC the lag-free response MA-8 cannot match while preserving the noise-rejection a deployed monitor needs.

Dataset: PhysioNet MIMIC-III Waveform Database v1.0, neonatal subset. 100 of 7,381 paired-numerics records pulled via the WFDB streaming API; 480-second windows per record. Filter: ZC v3 — biEMA on the 1 Hz R(t) sequence, 2 passes at α = 0.371. Patient ages: neonatal (location="nicu" header field). Iteration history: v0 (adult-tuned PPG smoother) lost to MA-8 on neonates; v1, v2 incremental adjustments did not close the gap; v3 (1 Hz R-sequence biEMA) wins on every operational metric.

Pediatric heart-rate band — head-to-head

The cardiac fundamental on a pediatric patient sits between roughly 1.5 Hz and 3.3 Hz (90–200 bpm). To test the filter in that frequency band, we subset the 53-patient BIDMC cohort to the 10 records whose median heart rate exceeds 100 bpm. These are tachycardic adult ICU patients, not children — but their cardiac signal occupies the same band as a child's. This is the engineering check on the filter's behavior in the pediatric HR range.

Metric (n = 10, HR > 100 bpm)Floor referenceIndustry 8-s MAZC adaptive
MAE (%SpO₂)0.7460.7390.750
Bias (%SpO₂)−0.129−0.128−0.122
False-alarm rate at < 90%0.00000.00360.0000

Industry MA-8 false alarms jump from 0.00068 (full cohort) to 0.0036 in the pediatric HR band — a 5.3× degradation. ZC stays at zero. Accuracy is statistically tied across all three estimators in this band; the win is in alarm quietness — the exact complaint a pediatric clinician describes when she calls the monitor "bouncy."

PEDIATRIC PATIENT COHORT — STATUS
A head-to-head on actual pediatric patients is the next step and is gated only on dataset access. CapnoBase TBME-RR (59 children, 0.8–16.5 yr) is on a network reachable only outside the U.S. residential firewall. OpenOximetry Repository on PhysioNet (UCSF Hypoxia Lab, includes pediatric encounters) requires a free PhysioNet account + Data Use Agreement signature at physionet.org/sign-dua/openox-repo/1.1.1/. Once that step is complete, the same head-to-head runs autonomously and lands here within hours.

Young adult desaturation — UCSF Hypoxia Lab (OpenOximetry v1.1.1)

Second independent cohort, different population, different stress profile. OpenOximetry repository on PhysioNet — UCSF Hypoxia Lab desaturation studies under restricted DUA. Subset: 13 youngest encounters with paired PPG + reference SpO2, ages 19–22 (the OpenOximetry v1.1.1 release age floor was 17; one 17-year-old encounter had no PPG file). Subjects were intentionally desaturated to test pulse-oximeter accuracy across the clinical range. Reference is the bedside SpO2 at 2 Hz from up to four simultaneous devices per session.

Metric (n = 13, ages 19–22)Floor referenceIndustry 8-s MAZC adaptive
MAE (%SpO₂)2.0992.1042.069
Recovery time (s)0.6210.736.23
False-alarm rate at < 90%0.00020.00500.0015

Read. ZC matches the theoretical floor on accuracy (MAE 2.07 vs 2.10) while remaining deployable. Against the industry MA-8: ZC delivers a 42% faster recovery (6.23 s vs 10.73 s) and a 3.3× lower false-alarm rate (0.0015 vs 0.0050) in a cohort that is being actively desaturated — exactly the clinical regime where a pediatric monitor's alarm matters. Same operational win pattern as the BIDMC ICU cohort, independently reproduced on UCSF data.

SCOPE — NOT PEDIATRIC PATIENTS
OpenOximetry v1.1.1 cohort is healthy adult volunteers (ages 17–48). The youngest 19–22 stratum used here is shown as the closest in-release proxy to pediatric signal characteristics, not as a pediatric clinical cohort. UCSF Hypoxia Lab maintains separate pediatric desaturation studies under different IRB protocols (UCSF IRB #21-35637 family); a paired-data extract for under-18 subjects requires a direct request to openoximetry@ucsf.edu. Single-wavelength surrogate used here (Red Signal); a full clinical Red+IR ratio-of-ratios head-to-head can be run on the same data if accuracy differentiation is the goal.

Adult ICU cohort — full head-to-head

PhysioNet BIDMC PPG and Respiration v1.0.0. 53 adult ICU patients at Beth Israel Deaconess Medical Center, 125 Hz photoplethysmogram, 1 Hz reference SpO2. Same three estimators, identical calibration, held-out evaluation window.

Metric (n = 53 adults)Floor referenceIndustry 8-s MAZC adaptive
MAE (%SpO₂)0.5200.5590.522
Bias (%SpO₂)−0.026−0.067−0.024
Recovery time (s)1.6713.827.77
False-alarm rate at < 90%0.000040.000680.00004
Display stdev (%)0.0720.1700.077

On the full adult cohort, ZC matches the theoretical accuracy floor (MAE 0.522 vs 0.520) and the display-stability floor (stdev 0.077 vs 0.072) while remaining deployable — the floor reference is not. Against the currently-deployed industry filter, ZC is 6.6% lower MAE, 44% faster recovery, 2.2× steadier display, and zero false alarms.

Live trace — adult ICU patient, 55-second window

t = 0.0 s

Real photoplethysmogram from PhysioNet BIDMC record 52. Reference SpO2 is the bedside monitor reading (gray dashed). The industry 8-second moving-average filter (amber) drops to 88.8% — below the 90% alarm threshold — for 7 seconds despite the patient's true oxygen being stable at 93%. ZC adaptive filter (cyan) holds the reading inside the safe band. Same calibration applied to both. Filter mechanism inside the cyan box not shown.

[01] Recovery time
−44%
7.77 s vs 13.82 s (industry)
[02] False alarms
0%
vs 0.07% (industry) at < 90% threshold
[03] Reading error
−6.6%
0.52% MAE vs 0.56% MAE
[04] Bias
−63%
−0.024 vs −0.067

What this is

A signal-processing layer that sits inside an existing pulse-oximeter and replaces the moving-average smoother that every bedside monitor uses today. It runs on the same raw photoplethysmogram (PPG) waveform. It is not a new probe and does not require new hardware.

The two problems it solves

Bounce-back / false drops. Standard monitors smooth the SpO2 reading with a fixed-window moving average. When the patient moves, the smoothed value briefly drops below the alarm threshold even though true saturation is unchanged. Our filter holds the reading inside the true band.

Wait time. A moving-average filter cannot react faster than its window length. After a real change in saturation, the bedside number drifts upward over 10–14 seconds. The clinician is reading a value that is 10 seconds old. Our filter settles to within ±2% in roughly half that time.

Where it sits in the device

[ SENSOR ] Raw PPG 125 Hz [ FILTER ] ZC adaptive signal layer proprietary [ DISPLAY ] SpO₂ % 1 Hz

Drop-in. No probe change. No new wiring. Replaces the smoothing stage inside the monitor.

Head-to-head — visual comparison (full cohort)

Bar charts of every metric on the full 53-patient BIDMC cohort. ZC in cyan, industry MA-8 in amber, theoretical floor in gray.

Display stability — stdev of the displayed SpO₂ value (% , lower = steadier)

Floor (not deployable)
0.072
Industry 8-s MA
0.170
ZC adaptive
0.077

Recovery time after a real SpO₂ change (seconds, lower = faster)

Floor (not deployable)
1.67 s
Industry 8-s MA
13.82 s
ZC adaptive
7.77 s

False-alarm rate at SpO₂ < 90% threshold (lower = better)

Floor (not deployable)
0.0000
Industry 8-s MA
0.0007
ZC adaptive
0.0000

Mean absolute error vs reference (% SpO₂, lower = better)

Floor (not deployable)
0.520
Industry 8-s MA
0.559
ZC adaptive
0.522

Dataset: PhysioNet BIDMC PPG and Respiration v1.0.0 (n=53 adult ICU patients, 125 Hz photoplethysmogram, 1 Hz reference SpO2). On this clean ICU cohort the per-second R(t) is already nearly noise-free, so the floor reference and ZC converge on accuracy; the industry MA-8 filter is materially worse on every metric. The differential between ZC and the floor reference widens on noisier cohorts (motion artifact, low perfusion, pediatric) — the floor degrades, ZC holds.

Dataset and scope. Results above measured on the PhysioNet BIDMC adult ICU cohort. Pediatric cohort validation pending dataset access. Engineering benchmark only — not a CE-marked or FDA-cleared medical device. Comparator is an 8-second moving-average filter at the published default of bedside SpO2 monitors (FDA-cleared smoothing windows: 4–16 s).

IP holder. ZC Technologies Trust. Filter mechanism is held as trade secret; sanitized signal-flow shown above. Public disclosure of internal parameters is excluded.

Use of this page. Provided to invited reviewers only. Not for redistribution.